Bland-Altman analysis answers the question "do two measurement methods agree?" It compares a new method against a gold standard or compares two methods on the same scale. The analysis does not test whether methods are identical, but whether differences between them are clinically acceptable and consistent. The APA 7 write-up reports bias (mean difference), its 95% confidence interval, 95% limits of agreement, and a test for proportional bias. This guide compares two blood pressure measurement methods on 50 participants.
What you need before you write a single word
An APA 7 write-up of Bland-Altman analysis needs:
- Sample size and a description of the participants or measurements.
- The two methods being compared and the measurement scale (e.g., mmHg, kg, seconds).
- Bias: the mean difference between methods and its standard deviation.
- 95% CI for the bias (computed using t distribution).
- Limits of agreement: mean difference ± 1.96 × SD of differences.
- 95% CIs for each limit (upper and lower).
- Proportional bias test: regression of differences on average values, with slope, t, df, and p.
- Clinical judgment: whether the limits of agreement are acceptable for clinical or scientific use.
The APA 7 format template
Two measurement methods were compared in [N] participants. Bland-Altman analysis revealed a bias of X.XX, 95% CI [X.XX, X.XX], indicating that [Method B] read [higher/lower] than [Method A] on average (SD of differences = X.XX). The 95% limits of agreement were between X.XX and X.XX, 95% CI [[X.XX, X.XX], [X.XX, X.XX]]. A test for proportional bias (regression of differences on average values) yielded slope = X.XX, t(df) = X.XX, p = [.XXX / < .001], indicating no evidence of proportional bias. Against a tolerance of ±10 mmHg set in advance, the limits of agreement lie inside it, but the upper limit's 95% CI (to 11.74 mmHg) does not. Within the limits, [X]% of differences were [clinically/scientifically acceptable / within [specify range]].
Want to see it written from data? Run this test on sample data in KyroStat: free, no file needed, and it does not use your free tests.
Formatting rules reviewers check:
- Bias and limits of agreement to 2 decimals.
- 95% CIs in square brackets, one CI per limit: upper LOA, 95% CI [X.XX, X.XX]; lower LOA, 95% CI [X.XX, X.XX].
- Report the regression slope for proportional bias, with t, degrees of freedom and p.
- State the measurement scale (e.g., mmHg, kg, days) so readers know the magnitude of agreement.
- Use "limits of agreement", not "agreement limits" or "limits of concordance".
- Clinical judgment: always comment on whether the limits are acceptable for practice. A ±5 mmHg LOA is excellent for BP; ±10 mmHg may not be.
- Italicize t, p, N, degrees of freedom.
A worked example
A hospital compared a new automated blood pressure device (Method 2) against a reference sphygmomanometer (Method 1) in 50 patients.
Descriptives:
- Method 1: M = 135.48 mmHg, SD = 13.46
- Method 2: M = 137.18 mmHg, SD = 14.07
Bias analysis:
- Bias (Method 2 - Method 1) = 1.69 mmHg, 95% CI [0.53, 2.86]
- t(49) = 2.92, p = .005: statistically significant, but only 1.69 mmHg on average
Limits of agreement:
- Lower LOA = 1.69 − 1.96 × (SD of differences) = −6.35 mmHg, 95% CI [−8.35, −4.34]
- Upper LOA = 1.69 + 1.96 × (SD of differences) = 9.73 mmHg, 95% CI [7.73, 11.74]
- Interpretation: 95% of differences are expected to fall between −6.35 and 9.73 mmHg
Proportional bias test:
- Regression slope = 0.05, t(48) = 1.07, p = .292
- No evidence of proportional bias: differences do not systematically increase with higher BP values
Agreement assessment:
- Mean bias: 1.69 mmHg (within the ±5 mmHg mean-difference value of AAMI/ISO 81060-2 Criterion 1)
- SD of differences: 4.10 mmHg (within standard of ≤8 mmHg)
- All 50 pairs (100%) fell within ±10 mmHg
- Decision: against a tolerance of ±10 mmHg set in advance, the limits of agreement (−6.35, 9.73) lie inside it, but the upper limit's 95% CI (to 11.74) does not, so agreement is acceptable only if ±10 mmHg is clinically acceptable for this use
Written up in APA 7:
Two blood pressure measurement methods were compared in 50 patients. The reference sphygmomanometer (Method 1, M = 135.48, SD = 13.46 mmHg) and the automated device (Method 2, M = 137.18, SD = 14.07 mmHg) were applied to each patient. Bland-Altman analysis revealed a bias of 1.69 mmHg, 95% CI [0.53, 2.86] (SD of differences = 4.10), indicating that the automated device reads higher than the reference on average. The 95% limits of agreement were between −6.35 and 9.73 mmHg, 95% CI [−8.35, −4.34] to [7.73, 11.74]. The mean bias and the SD of differences fell within the values of AAMI/ISO 81060-2 Criterion 1 (mean difference within ±5 mmHg, SD at most 8 mmHg); with 50 patients rather than the 85 the standard requires, this is an illustration, not a device validation. A test for proportional bias (regression of differences on average values) yielded slope = 0.05, t(48) = 1.07, p = .292, indicating no evidence of proportional bias.
Understanding the Bland-Altman components
Bias (mean difference). This is the systematic difference: does one method consistently read higher or lower? Statistical significance does not equal clinical significance; a 1.69 mmHg mean difference may be real but negligible. The 95% CI shows the range of plausible population biases.
Limits of agreement. These are ±1.96 × SD of the differences. The range defines where 95% of individual differences fall. Wide LOA indicate disagreement too large for interchangeable use; narrow LOA indicate close agreement.
95% CIs for the limits. These show how precisely the LOA estimates are known. With small samples (N < 30), the CIs widen due to limited sample size.
Proportional bias. Does the difference between methods depend on the measurement value? A positive slope means larger values show larger differences. If this test is significant, agreement is not constant across the measurement range.
Bland-Altman plot
A Bland-Altman plot graphs each pair's difference on the y-axis versus the average on the x-axis. The plot shows:
- A horizontal line at the bias (mean difference)
- Horizontal lines at the upper and lower limits of agreement
- Each pair as a point; outliers are obvious
Difference (Method 2 - Method 1)
↑
10 | · Upper LOA = 9.73
| · · · ·
5 |· · ·
| · · ·
|--·---·-- Bias = 1.69
0 | · ·
| · ·
-5 |·
| Lower LOA = −6.35
-10 |
|_____________________→ Average (Method 1 + Method 2)/2
100 120 140 160
Reviewers often want this plot to see outliers and check assumptions.
Mistakes reviewers catch
- Reporting only the bias, not the limits of agreement. The limits show where individual differences lie, not just the average difference. Both matter.
- Missing 95% CIs for the limits. These show the precision of the LOA estimates; small samples have wide CIs, indicating uncertainty.
- Testing whether bias is statistically significant, not clinically significant. With large N, a 0.5 mmHg mean difference will be statistically significant but clinically irrelevant. Focus on the limits and clinical judgment.
- Ignoring proportional bias. If agreement is better at lower values and worse at higher values (or vice versa), the methods are not interchangeable across the full range.
- Confusing Bland-Altman with correlation. Two methods can be highly correlated but have wide limits of agreement. Correlation tests association; Bland-Altman tests agreement.
- Not stating a clinical threshold or decision. End with "these limits are [acceptable / too wide] for clinical use because [reason]." Leave no ambiguity.
- Using parametric CI methods with skewed differences. If differences are non-normal, consider a log transformation or non-parametric bootstrap CIs.
Why Bland-Altman instead of correlation or paired t-test?
| Goal | Statistic | Why not others |
|---|---|---|
| Test if means differ | t-test | Only tests average; ignores individual variability |
| Test if linearly related | Correlation | Only tests association; two methods can correlate perfectly but disagree |
| Assess agreement range | Bland-Altman limits | Shows individual differences and clinical acceptability |
Bland-Altman analysis is specifically designed for method agreement. Use it when comparing two measurement techniques.
Let KyroStat do the analysis for you
KyroStat computes the bias, its 95% CI, the limits of agreement with their 95% CIs, the proportional bias test, and reports the agreement range and clinical interpretation. Upload two columns of measurements and get the full APA 7 write-up in seconds.
Frequently asked questions
What if my sample is small (N < 30)? Bland-Altman still works, but the 95% CIs for the limits are wide, reflecting uncertainty. Report them fully; they show the limits may be less precise than a larger study would reveal.
Should I report both upper and lower limits of agreement? Yes, always. They are symmetric around the bias by construction (bias ± 1.96 SD); they are only asymmetric after a log transform or in regression-based limits.
Can I use Bland-Altman to compare more than two methods? Run Bland-Altman for each pair separately. For 3 methods (A, B, C), report A vs B, A vs C, and B vs C. Conclusions should be consistent (if A agrees with B and B agrees with C, then A and C should roughly agree).
What if the differences are not normally distributed? Bland-Altman assumes normality of differences for the LOA. If differences are very skewed, consider log-transforming the original measurements or using bootstrap methods to compute CIs. Report any transformation.
How do I know if the limits are "acceptable"? This is a clinical or scientific judgment, not statistical. For blood pressure, ±5 mmHg might be excellent; for weight, ±1 kg. Know your field's standards and state your threshold clearly.
What is proportional bias, and why test for it? Proportional bias means the difference between methods changes with the measurement value. If the automated BP device reads 1 mmHg higher at BP = 110 but 5 mmHg higher at BP = 180, agreement is not constant. This matters if the methods will be used across a wide range.
Can I use Bland-Altman on repeated measures within individuals? Bland-Altman compares two methods on the same scale. If you have repeated measures over time within individuals, follow Bland and Altman (1999), who give methods for several pairs per person that account for within-subject variation; pooling all pairs as if they were independent understates the standard errors. Running a separate analysis per time point is the simple alternative. Averaging measurements per individual removes within-subject variance and may hide disagreement at specific time points.
Related guides
- How to report correlations in APA 7: testing association, not agreement.
- How to report paired t-tests in APA 7: testing mean difference, one component of agreement.
- How to report linear regression in APA 7: the proportional bias test is a simple linear regression.
- How to report confidence intervals in APA 7: CIs around the bias and limits.
- Which statistical test should I use?: picking the right analysis.